Specifications
CE ISO Approved
Peristaltic finger pump
Purge rate and KVO rate
Boluse Mode, Time mode
8 Level alarm setting
Infusion Pump (JAS-1300)
Infusion Mode: Drip-rate and volumetric
Pumping Mechanism: Peristaltic finger pump
Optional Infuse Rate Unit: d/min, mL/h and volume/time
IV Sets: Standard IV sets (20d/mL and 60d/mL) with diameter 3.4 ~ 4.5mm
Flow Rate Range: 1.0 ~ 1000 mL/h
Drip Rate Range: 20 d/mL IV set: 1.0 ~ 333.3 d/min; 60 d/mL IV set: 1.0 ~ 150.0 d/min
Increment: 0.1 mL/h
Flow Rate Accuracy: ≤ ± 5%(after calibration)
Drip Rate Accuracy: ≤ ± 1.5%
Volume Limit: 1 ~9999 mL
Volume Infused Display: 0 ~65000 mL
Purge Rate: 600 mL/h
KVO Rate: Adjustable 1 ~ 5 mL/h
Bolus Mode: Available for Bolus mode, flow rate adjustable.
Occlusion Pressure: Pressure detection device with 8 sensitivities setting, 40 ~ 140KPa.
Air-in-line Detection: Air bubble detected by ultrasonic and the device with 8 sensitivities setting, 50 ~ 1000μL.
Alarm Volume: 8levels setting
IV Set Management and Flow Rate Calibration:
IV sets can be added, stored and deleted in memory, and easily used after calibration.
Self-check: Self-check when start up and during infusion
Audio and Visual Alarms:
Air-in-line, Door open, Occlusion, Infusion finished, Pause overtime, Leakage, Empty, Low battery, Battery damaged, Abnormity of drop detection, infusion, motor, sound, drip sensor, pressure sensor, ultrasonic sensor, and AC power supply.
Display: Blue LCD with high brightness
Communication Port: RS-232 (for nurse call)
Power Source: AC: 100V ~ 240V, 50Hz/60Hz, ≤ 25VA
Rechargeable Ni-MH battery: DC12V, 2000mAh, life: ≥ 4 hours at 25mL/h, remaining battery displayed.
Operating Conditions: Temperature: 5°C ~ 40°C
Relative humidity: 20% ~ 90%
Atmospheric pressure: 70 ~ 106KPa
Dimensions and N.W.: 150mmx 150mm x 235mm; 2.3kg
Application: Intravenous infusion for various liquid medicines, such as vascular expansion solution, anticancer drug, accelerative procreates drug, liquid nutrient, etc.
Applicable for micro and rapid infusion.