High quality CS996 High-frequency Carbon Sulfur Element Analyzer Unit/high- frequency infrared carbon sulfur analyzer
USD $14,000 - $18,000 /Set
Min.Order:1 Set
Wuxi Jiebo Electrical Technology Co., Ltd.
CS996 Sulfur Element Analyzer Unit
Wuxi Jiebo Electrical Technology Co.,Ltd is a manufacturer on analytical instruments, enjoying a good reputation and feedback from our customers domestic and overseas. We provide main machines, as following:
Main Parameter
Measure Range:
Sample weight:
standard 0.5g
Analysis Error:
Carbon: accord with ISO9556 standard
Sulfur: accord with ISO4935 standard
Analysis Precision:
Carbon: RSD<0.5%
Sulfur: RSD<1%
Analysis pool:
Carbon: low carbon pool and high carbon pool
Sulfur: if the customer request,the high sulfur pool could be added
Time of analysis:
25 to 60 seconds,could be adjusted. Around 35 seconds usually
Sensitivity(The minimum readings):
C/S: 0.1ppm
Electronic scale:
Precision of read: 0.0001g
Working Environment:
Room temp: 10 to 30 degree C,relative Humidity less than 75%
Infrared Detection Unit
Circuit Design: | The modularized design with double CPU for the whole frame is divided into the upper and lower levels while the electronic circuits are highly integrated stably and reliably;<0{>in the meantime the multilevel hidden-type isolated circuit is employed so that it is free from HF interference. |
Power: | The linear close-coupled modular power block ensures stable output without fault. |
Light source: | The tailor-made new model platinum infrared light source has lasting caloricity and efficient spectral property. |
Analysis cell: | The gilded carbon & sulfur analysis cell is fit with the high precision TSC infrared detectorl.It is also possible to have the double-carbon cell or the double- carbon & double-sulfur cell fit out at the client’s request. |
Motor: | The special-purpose aviation modulation motor is applied with satisfactory thermal stability whose life expectancy exceeds 100,000 hours. |
Installation of the instrument
Preparation before the installation
Note: The user should make appropriate working preparation before the arrival of company installation staff.
1. Analyzing room: be far away from the erosion gas such as acid, alkali, dust, vibration, testing disturbance.
2. Analyzing room size: >3×3m;
3. Working condition: Room temperature: 10-30°C, Relative humidity: <75%;
4. Power supply: well earthing is required, Voltage: AC220V±5%, Frequency: 50Hz±2%, No disturbance from harmonic wave
5. Voltage stabilizing equipment: Power: 5kW, Voltage stabilizing precision<2%;
Note: If there is disturbance from intermediate frequency furnace in the work power supply, the equipment should be equipped with AC Regulated Power Supply.
6. Gas: Oxygen: purity≥99.5%
7. Power gas: Nitrogen or Compressed air (Purified: without water, oil and contamination)
8. Tool: Box opening tool and regular tools.
If interested in,pls don't hesitate to send your enquiry,and welcome to visit our company in anytime.