Mean well RPS-160-48 160W 48V Medical power supply 160W 48V medical power supply
USD $1 - $100 /Piece
Min.Order:1 Piece
Hangzhou Nuohan Electronics Co., Ltd.
Mean well 20w 24v single output power supply/20W Single Output Encapsulated Type/power supplies 20w
Mean well 15w 5v power supply/15W Single Output Encapsulated Type
Mean well 15w 3.3v power supply/15W Single Output Encapsulated Type/15w switching power supplies
Meanwell 5v 24v 12v -12v Switching Power Supply/200W Medical Switching Power Supply
MODEL | RPS-160-48 | |
OUTPUT | DC VOLTAGE | 48V |
RATED CURRENT (20.5CFM) | 3.25A | |
CURRENT RANGE (convection) | 0 ~ 2.3A | |
CURRENT RANGE (20.5CFM) | 0 ~ 3.25A | |
RATED POWER (convection) Note.7 | 113.4W | |
RATED POWER (20.5CFM) Note.8 | 161W | |
RIPPLE & NOISE (max.) Note.2 | 250mVp-p | |
VOLTAGE ADJ. RANGE | CH1: 43.2~ 52.8V | |
VOLTAGE TOLERANCE Note.3 | ±2.0% | |
LINE REGULATION | ±0.5% | |
LOAD REGULATION | ±1.0% | |
SETUP, RISE TIME | 1800ms, 30ms/230VAC 3500ms, 30ms/115VAC at full load | |
HOLD UP TIME (Typ.) | 16ms/230VAC/115VAC at full load | |
INPUT | VOLTAGE RANGE Note.6 | 90 ~ 264VAC 127 ~ 370VDC |
FREQUENCY RANGE | 47 ~ 63Hz | |
POWER FACTOR (Typ.) | PF>0.93/230VAC PF>0.98/115VAC at full load | |
EFFICIENCY (Typ.) | 88% | |
AC CURRENT (Typ.) | 2A/115VAC 1.1A/230VAC | |
INRUSH CURRENT (Typ.) | COLD START 35A/115VAC 70A/230VAC | |
LEAKAGE CURRENT Note.9 | Earth leakage current<200uA / 264VAC, Touch current < 100uA/264VAC | |
PROTECTION | OVER LOAD | 105 ~ 135% rated output power |
Protection type : Hiccup mode, recovers automatically after fault condition is removed | ||
OVER VOLTAGE | 55.2 ~ 64.8V | |
Protection type : Shut down o/p voltage, re-power to recover | ||
OVER TEMPERATURE | 110°C5V),105°C(12V,15V,24V,48V) (TSW1 : detect on heatsink of power transistor) | |
95°C(5V),90°C(12V,15V,24V,48V) (TSW2 : detect on heatsink of power transistor) | ||
Protection type : (TSW1)Shut down o/p voltage, recovers automatically after temperature goes down | ||
Protection type : (TSW2)Shut down o/p voltage, re-power on to recover | ||
FUNCTION | 5V STANDBY (G model) | 5VSB : 5V at 0.6A without fan, 0.8A with fan 20.5CFM ; tolerance ±2%, ripple : 50mVp-p(max.) |
PS-ON INPUT SIGNAL (G model) | Power on: PS-ON = "Hi" or " > 2 ~ 5V" ; Power off: PS-ON = "Low" or " < 0 ~ 0.5V" | |
POWER GOOD / POWER FAIL | 500ms>PG>10ms PF>1ms | |
REMOTE SENSE | 5 ~ 15V | |
ENVIRONMENT | WORKING TEMP. | -20~+70°C (Refer to derating curve) |
WORKING HUMIDITY | 20 ~ 90% RH non-condensing | |
STORAGE TEMP., HUMIDITY | -40~+85°C, 10~95%RH | |
TEMP. COEFFICIENT | ±0.03%/°C(0~50°C) | |
VIBRATION | 10 ~ 500Hz, 2G 10min./1cycle, Period for 60min.eachalong X, Y, Z axes | |
SAFETY&EMC (Note 4) | SAFETY STANDARDS | ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved |
WITHSTAND VOLTAGE | I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC | |
ISOLATION RESISTANCE | I/P-O/P, I/P-FG, O/P-FG:100M Ohms/500VDC/25°C/70% RH | |
EMCEMISSION | Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B, EN61000-3-2,-3 | |
EMC IMMUNITY | Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A | |
OTHERS | MTBF | 230.5K hrs min MIL-HDBK-217F (25°C) |
DIMENSION | 127*76.2*34.6mm (L*W*H) | |
PACKING | 0.32Kg; 36pcs/12.5Kg/0.79CUFT | |
NOTE | 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25°C of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor. 3. Tolerance: includes set up tolerance, line regulation and load regulation. 4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meetsEMC directives. 5. HS1,HS2 & HS3 can not be shorted. 6. Derating may be needed under low input voltages. Please check the derating curve for more details. 7. The rated power includes 5Vsb at 0.6A. 8. The rated power includes 5Vsb at 0.8A. 9. Touch current was measured from primary input to DC output. 10. Suitable for BF application with appropriate system consideration. |