high sensitivity LF-1200 air and water disinfector/ozone air fresheners/ozone sensor
Specifications:
The ozone gas alarm is intelligent instrument with a multifunctional, adopts advanced signal processing techniques, combination of hardware and software to realize the ozone measurements of high performance and low cost, the ozone sensor adopts imported HMOS ozone semiconductor devices, high sensitivity, rapid response, high precision, long service life, is very suitable for ozone leakage monitoring the alarm
Technical parameters:
Measurement object | Ozone in the air |
Sensor Type | Heating type metal oxide semiconductor sensor or an electrochemical sensor |
Measuring range | 0-1ppm (resolution 0.001ppm); 0-100ppm (Resolution 0.001ppm, 0.01ppm) |
Measurement accuracy | 0.001ppm;0.01ppm;±3%FS |
Analog Output | Optional 0-5V or 4-20mA (9 ADC) |
Digital output | RS485 |
Power supply | 12V DC (standard) (24V DC optional) |
Semiconductor sensor accuracy depends on the heating time is recommended first energized more than 24 hours to ensure accurate readings, after a short period of time useless, powered up again to warm up for at least 10 minutes, more than 1 week is recommended warm-up for 24 hours.
Sensor Interface | 1 | 2 | 3 | 4 | |
5V+ | HEAT | SIGN | GND | ||
Output Interface | 1 | 2 | 3 | 4 | 5 | 6 | 7 | |
Relay output | 4~20MA+ | 4~20MA- | RS485A | RS485B | NOP | |||
Power connector: 5.5 x 2.1 mm DC power connector
Sensor Interface: 4-core plug (Ø12)
Output Interface: core plug (Ø12)
Relay Output: normally open type contacts largest AC250V, 10A
Analog output: 0-5V and 4-20mA can not be used at the same time need to be specified when ordering, 0-5V output is non-positive and negative short-circuited
Show: four red LED display;
Run: run light, normal working hours flashes per second; ALARM: The alarm indicator, the ozone concentration is greater than the set alarm concentration Always, normally open relay closes.
Alarm point for the method:
1: Press and hold the SET key for more than 3 seconds, running lights and alarm indicator light, into the set, displays the current alarm value;
2: Press : alarm value increases, the value decreases ,Press alarm; Press and can be quickly adjusted;
3: adjustment is completed, press the SET button to return to run the interface, run the light is blinking;
Fault information:
1: ALARM light: ozone concentration values greater than the set alarm value, the concentration of ozone gas automatic recovery
2: ERR: sensor failure, check the connection is normal, or contact the manufacturer to replace the sensor
3: Display OVER: ozone concentrations exceeding the sensor range