high precision CL7685 ozone purifier machine/greenhouse air ozonator/auto-manual ozone machine
USD $1 - $4,000 /Unit
Min.Order:1 Unit
Shenzhen Long Life O3 Tech Co., Ltd.
sensitive LF-UVO3-1500 sterilizer water/ozone sensor/indoor ozonizer for air treatment
high sensitivity LF-1200 air and water disinfector/ozone air fresheners/ozone sensor
good quality LF-MC500 odor remover ozone/ozone water filter/wall mounted ozone analyzer
Ozone tube 1/2/10/15/30/50 Gram/ozone producer/ozone generator kits
high precision CL7685 ozone purifier machine/greenhouse air ozonator/auto-manual ozone machine
This series of controllers with built-in microprocessor forms an advanced system of measurement and regulation of pH - ORP - Conductivity - Dissolved Oxygen - Chlorine in industrial processes.
1.1 FUNCTIONAL SPECIFICATIONS
Input The instrument accepts input from potentiostatic sensor for Active Chlorine, Chlorine Dioxide and D.Ozone measuring. A second input is provided for 3 wires Pt100 Temperature sensor. Temperature compensation The unit is supplied with manual or automatic Temperature compensation and Temperature information may be displayed on LCD. The instrument detects the absence or malfunctioning of the Temperature sensor and automatically switches to manual operation mode. Measuring ranges The unit provides an input range which may be selected from 0/1.999 or 0/19.99 PPM. Auto ranging function may be activated for the measuring range 0/1.999 PPM. Auto ranging allows the operator to calibrate the unit in the low range, against a high concentration of standard solution.
Alarm relay
The unit contains a SPST relay designated as an alarm relay.
Cleaning function
The unit contains a SPST relay designed for the auto clean cycle. It's possible manually activate the cleaner to test its operation. The operator can change the frequency of cleaning. The configuration routine selects: - auto/manual operation - the cleaning time - the holding time.
Integrated ozone concentration sensor structure
This paragraph ozone concentration sensor and display by the host header of two parts, the structure as shown below:
1 Ozone gas out of the air intake (gas inlet, outlet diameter 4mm, diameter 6mm.)
2 Ozone gas out of the air intake (gas inlet, outlet diameter 4mm, diameter 6mm.)
3 Power and signal input and output terminals
Its power and signal input and output terminals are defined as follows:
Needle angle | 1 | 2 |
input | 24V+ | 24V- |
| White | Black |
Needle angle | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 |
output | 5V+ | 5V- | 485+ | 485- | Zeroing | Zeroing | 4-20mA+ | 4-20mA- |
| Brown | Red | Orange | Yellow | Green | Blue | Purple | Gray |
Ozone sensor display header
1.485 display header 2. Signal input terminals
3.front panel 4. Display window
5.Zeroing button 6.Zeroing button
The signal input terminals are defined as follows:
Needle angle | 1 | 2 | 3 | 4 | 5 | 6 |
input | 5V+ | 5V- | 485+ | 485- | Zeroing | Zeroing |
| Brown | Red | Orange | Yellow | Green | Blue |
Main Product Picture
Ozone Generator Analyzer/Monitor/Tester
http://longlifeozone.en.china.cn/product/60252745768-221802020/High_quality_LF_SP015G_Medical_ozone_generator_ozone_tube.html
Factory
Application
Certificate