10.4 '' TFT Display Screen Multifunctional Anesthesia Unit/Anesthesia Machine Price MSLGA13-4
USD $100 - $4,000 /Set
Min.Order:1 Set
Guangzhou Medsinglong Medical Equipment Co., Ltd.
10.4 '' TFT Display Screen Multifunctional Anesthesia Unit/Anesthesia Machine Price MSLGA13-4
General introduction of anesthesia ventilator
This Anesthesia Machine is equipped with an accurate dedicated anaesthetic vaporizer and a safety device for preventing cyanosis and necessary alarm system. During the anesthesia, the patient’s respiratory functions can be managed by using a microcomputer controlled pneumatic electrically controlled synchronizing anesthesia respirator. Each connection part of the whole machine is a standard interface. A highly efficient and large volume soda lime absorber can minimize the re-inhalation of carbon dioxide by the patient.
Technical Parameter
1. Operating mode: circulatory closed, semi-closed and semi-open.
2. Gas requirement: medical oxygen and nitrous oxide with a pressure ranging from 0.3 MPa to 0.5 MPa.
3.The maximum error of the pressure testing device should not exceed ± (4% of the full scale reading + 8% of the actual reading).
4. For oxygen and nitrous oxide, a special pressure regulator with a safety valve will be provided. The exhaust pressure of the safety valve should not be higher than 6 kPa.
5. Indication range of the oxygen, nitrous oxide and air flow meter: 0.1 L/min ~ 10 L/min.
When the flow rate ranges from 10% of the full scale value to 100%, the scale accuracy should be within ±10% of the indicated value.
6. The flow meter is equipped with an oxygen-nitrous oxide proportional control device.
When the oxygen concentration in the mixed gas N2O/O2 conveyed by the anesthesia
machine is no lower than 20% (V/V) or the FiO2 is lower than 20%, the machine will emit an
alarm.
7. When the oxygen pressure of the anesthesia machine is 0.20MPa±0.05MPa, the machine raises a low gas source pressure alarm which is a high priority alarm, and shuts off the nitrous oxide conveyed to the common gas outlet.
8. Oxygen Flush:25~75 L/Min;
9. Anesthetic gas concentration adjustment range of vaporizer: 0~5%, relative error ±20 %.
10. The exhaust pressure of the safety valve of the respiratory circuit is no higher than 6 kPa.
11. Anesthesia ventilator
* Respiration Mode:IPPV, SIPPV, IMV, SIMV, PEEP, Manu, Sigh and so on.
* Respiration Frequency:1~99bpm
* I/E Ratio:4:1~1:10
* Tidal Volume:0~1500mL
* IMV frequency:1~12bpm
* Ptr:-10~10 hPa;
* Controlled ventilation and assisted ventilation switching time: 6s
* Inspiratory Plateau:0~1S;
* PEEP:1.5~10 hPa;
* O2 Concentration:21%~100%
* Sigh: one deep breath per 80 controlled respirations, the inspiration time is 1.5 times of the setting point and the permissible error is -10% or 15%.
* Maximum safety pressure: ≤ 12.5 kPa.
* Pressure limit range: 1~6 kPa
Detailed Pictures
1.Manual/Maneuvering changeover valve 10.APL safety valve
2.Common gas inlet 11.Air bellow
3.Gas storage bag connection port 12.Driving gas inlet
4.Carbon dioxide absorber 13.Pressure signal
5.Exhalation valve 14.Oxygen concentration sensor interface
6.Exhalation screwed pipe port 15.Waste gas exhaust port
7.Inhalation screwed pipe port 16.Connection with 17 through screwed pipe
8.Inhalation valve 17.Connection with 16 through screwed pipe
9.Pressure gauge
We can offer you various shipping choices, guranteeing safe and fast delivery
Found in 2004, Located in Guangzhou City, with 20,000M2 Industrial Park & over
300 employees, Guangzhou MedSingLong Medical Equipment is one of the
biggest Factory Based Professional medical equipment Supplier in China.
Under the strict Management of ISO9001: 2008 & CE. MSL had been developed
into a Big Scaled Manufacture in China.
We believe only you succeed, we will succeed. That's why our slogan is :
(MSL)Medical Supplier for Long-term
For further communication, you can either send me inquiries or via these tools. :)