Specifications
1.Microprocessor based servo controlled temperature system2.Alarm functions
3. Heater head and bed angle adjustable
4. X-ray
Specifications
baby infant warmer
1 CE baby infant warmer
2 safe baby infant warmer
3 medical baby infant warmer
4 Hot infant warmer
Model: baby infant warmer BN-100 Standard
Baby infant warmer BN-100 standard brief introduction:
1. Microprocessor based servo controlled temperature system.
2. The microprocessor-controlled unit allows for three settings: Pre-warm, manual or automatic.
3. Alarm functions for safety includes: Power failure, temperature deviation, Temperature sensor failure, over temperature.
4. APGAR timer to record therapy time
5. Convenient X-ray cassette under infant bed
6. Heater head can be adjusted +90 degree horizontally
7. Tilt angle of infant bed is adjustable
8. Observe light angle is adjustable
9. Transparent protector folded
10. Adopts anti-explosion micro crystal quartz infrared radiation tube for its radiation source, the heating is quick and life-span is long
11. Alarms are audible and visual
12. A second thermal cut-out function for more safety
13. Skin temperature sensor failure protect function avoid over temperature
14. Temperature deviation can be amended in front panel directly
Baby infant warmer BN-100 (standard) standard configuration
Main body (including the radiant source, control system, infant bed, bracket), X-ray cassette, Observe light, I. V. Pole, skin temperature sensor, tray, mattress, transparent protector, castors.
Baby infant warmer BN-100 (standard) optional parts: Drawer, RS-232(485) connecter
Baby infant warmer BN-100 (standard) Technical data:
1. Power supply: 110/220V 60/50Hz
2. Power input: ≤ 1200VA
3. Operating condition
An ambient Temperature: 18°C-30°C
An ambient Relative Humidity: 30--75 percent RH
Atmospheric Pressure: 700~1060 hPa
An ambient Air Velocity <0.3m/s
4. Temperature control: 32~38°C
5. Temperature control accuracy: ≤ 0.5°C
6. Mattress temperature uniformity: <25°C
7. Skin temperature sensor accuracy: + /-0.3°C
8. Warm-up time: ≤ 30min
9. Infant bed tilt angle: + 10degree
10. APGAR timer range: ≤ 24h
11. Heater head horizontal: 90degree;
12. Mattress size: 67cm X 54cm
13. Distance from heater to mattress: 80cm
Baby infant warmer BN-100 (standard) Transport and storage
Environment Temperature: -40~ +55°C
Environment Relative Humidity: ≤ 95percert
Atmospheric Pressure: 500~1060hPa