Medical X-Ray Protection Lead Glass CMW-141416
Negotiable /Piece
Min.Order:5 Pieces
Function: Lead Glass is applied for X ray ,γ ray radiation shielding.
Field: It is widely used for radiation protection in field of medical diagnose,treatment activity observation windows in x ray room and isotope partial protection.
Size: 356x356x8mm
Lead equivalent: 1.6mmpb
Voltage: x ray tube peak voltage between 100kv to 300kv
Lead Glass Technical Data
Model | Density (g/cm3) | Lead Equivalent (mmpb/mm) | Light transmission | X ray tube peak voltage | |
X ray | γ ray (Gamma) | ||||
ZF2 | 4.12 | 0.21 | 0.36 | 86% | 120 |
ZF3 | 4.46 | 0.24 | 0.38 | 86% | 200 |