Product Details

Specifications

1.wavelength coverage:340-1000nn
2.spectral bandwidth:4nm
3.wavelength repeatability:0.6%T

Description:

1). The instrument work, must be preheated can be measured only 20 minutes
2).Wavelength selection, turn the wavelength window display, adjust to the need of test wavelength
3). Set up the test mode, press the function key, can switch test mode (A: Abs absorbance, T: Trans light transmittance, C: Conc: concentration
4). After get the test result, press the "print" button to print the results
5). Zero (0% T)
Put the occulter into the sample frame, and pull the rack lever makes it into the light path, and then 0% instrument of "T" button, can complete the zero.
Note:
A: zero time road must be put into gear, the zero block, otherwise unable to complete zero, showing "E01"
B: don't open the sample chamber cover
C: zero in transmittance mode according to the "00.0", in absorbance mode shows "1 -", which is beyond the scope display
6). Adjust the (100% T)
A: Can't block zero block into the light path, otherwise cannot be the best, show "E02"
B: don't open the sample chamber cover hundreds
C: in absorbance mode is shown as "000", in transmittance mode is shown as "100.0"
7). The absorbance measurement
1. According to the "key" to switch to the light transmittance measurement model
2. Adjust the test wavelength
3. Placement window, press the "T" 0% to zero
4. Click "function key" to switch to the absorbance measurement model
5. Put reference sample, press the "T" 100% and instrument shows "000".
6. Placement test samples, read the test data
7. Measured values more than 1.999, shows "1 -" which is beyond the scope display
8. When the measured value is lower than 1.999, showing "1 -" which is beyond the scope display
8). Light transmittance test
1. According to the "key" to switch to the light transmittance measurement model
2. Adjust the test wavelength
3. Placement window, press the "T" 0% to zero
4. Placement reference sample, press the "T" 100% best, instrument display "100.0"
5. Placement test samples, read the test data
6. Measured values more than 199.9, shows "1 -" which is beyond the scope display
9). Standard concentration measurement

 

 

 

 

 

Contact Supplier

Ms. Tina Li Manager Chat Now
Telephone
86-755-28115540
Mobile
18923861750
Fax
86-755-29822470
Address
B602, Floor 6, Huibaojiang Building, Minzhi Avenue, Minzhi Street., Bao'an District Shenzhen, Guangdong, China (Mainland) Shenzhen,Guangdong

You May Like

New Products

Recommended Products

View More View Less

Find Similar Products By Category

Measurement & Analysis Instruments > Other Measuring & Analysing Instruments
Categories
Hi~