PXSJ-216F Benchtop Ion meter LCD pX/Ion/mV (ORP)/temp Accuracy grade:0.001
Negotiable /Set
Min.Order:1 Set
Shanghai Yanhe Instrument Equipment Co., Ltd.
Functions:
1.It can measure pH/pX, ion concentration, potential and temperature.
2.It adopts dot-matrix LCD and light touch keys. Power-failure protection function.
3.It can recognize several pH buffers which meet with GB, NIST and DIN standard. Up to 5 points can be calibrated. It has automatic temperature compensation function.
4.It supports the switch between various concentration units.
5.It supports several ion modes such as H+, Ag+, Na+, K+, NH4+, Cl-, F-, NO3-, BF4-, CN-, Cu2+, Pb2+, and Ca2+ . The user can also set up specific ion mode.
6.It provides several ion concentration measurements including direct reading of concentration, standard addition mode, sample addition mode and GRAN mode.
7.It supports three measuring modes including continuous mode, timed reading mode and auto-lock reading mode which meet with different requirement.
8.It can delete, view and print data. Up to 200 sets of pH data, 100 sets of six ion data and 100 sets of concentration data can be saved.
9.It has USB interface. If matched with specific communication software, the meter can be connected with computer.
Specifications:
Model | PHSJ-4F | PHSJ-3F | PXSJ-216F |
Parameters | pH /mV(ORP)/℃ | pX/ Ion concentration /mV (ORP)/ ℃ | |
Range | pH: -2.000~20.000 mV: ±1999.9 Temp: -5.0~110.0 |
pH: -2.00~20.00 mV: ±1999 Temp: -5.0~135.0 | pX: -2.000~20.000 mV: ±1999.99 Ion concentration: (0~1999.0), unit ug/L, mg/L, g/L, mol/L, mmol/L Temp: -5.0~110.0℃ |
Resolution | pH: 0.1/0.01/0.001 mV: 0.1 Temp: 0.1℃ | pH: 0.01 mV: 1 Temp: 0.1℃ | pX: 0.001 pH/pX mV: 0.1 Temp: 0.1℃ |
Accuracy | pH: (±0.002pH±1 bit)/3h mV: ±0.03%FS Temp: ±0.2℃±1 bit | pH: (±0.01pH±1 bit)/3h mV: ±0.1%FS Temp: ±0.2℃±1 bit | pX: ±0.002pH/pX mV: ±0.03%FS Ion concentration: ±0.5% Temp: ±0.2℃ |
Ion Mode | — | Provide several general ion modes, such as: H+, Ag+, Na+, K+, NH4+, Cl-, F-, NO3-, BF4-, CN-, Cu2+, Pb2+, Ca2+, etc… and support customer ion mode | |
Temp Compensation | Manual/Auto (-5.0~110.0)℃ | Manual/Auto (-5.0~135.0)℃ | Manual/Auto: 0.0~60.0℃ |
Calibration | May select 5 standard buffer solution for three-point calibration | Five-point calibration | |
GLP standard | Yes | ||
Data storage | 500 sets of data | 1400 sets of data | |
Buffer Recognition | Automatic: Five Buffers | Automatic: GB,NIST,DIN | |
View, Delete&Print | Yes | ||
Interface | USB | ||
Power | Universal power adapter (9V DC, 800mA) | ||
Dimension&Weight | 280×215×92mm, 1kg | ||
Standard Configuration | One-in-three pH Combination Electrode; Standard Buffer | One-in-three pH Combination Electrode; Standard Buffer | PF-1-01 Fluoride ISE; 232-01 Reference Electrode; T-818-B-6 ATC Probe; JB-1A Stirrer |