Main technical parameters
1. Number of treatment channels
1. 1 Every one of the 8 treatment channels can work and adjust independently.
1. 2 Two roads magnetic field output, two roads top nuclear power electrical stimulation output, four ways body electrical stimulation output.
2. Output index
2. 1 The indicator lamp and a sound will prompt the running of every channel.
2. 2 The digital clock start counting down when every channel runs.
3. Magnetic field output
3. 1 Magnetic displacement of treatment apparatus
Strong gear: 14mT±4mT;
Weak gear: 7mT±2mT.
3. 2 Treatment time
3. 2. 1 There are two gears, 20 min and 30 min, of treatment time to be selected from before starting treatment.
3. 2. 2 Treatment time error is ±1min.
3. 2. 3 The initial default treatment time of the main machine in operation is 20 min.
3. 3 Magnet therapy model
3. 3. 1 Variable frequency model: (F1,F2,F3,F4,F5,F6) the frequency is automatically switched over every 10s.
3. 3. 2 Fixed frequency model: choose one frequency from F1-F6 before the treatment starts.
3. 3. 3 It starts with variable frequency model.
3. 4 Magnet therapy frequency
3. 4. 1 The treatment has 6 frequency stages: (5,10,20,30,40,50)HZ.
3. 4. 2 It starts with the frequency of F1(5)HZ.
3. 4. 3 Autocycle variable frequencies within F1—F6 is usually adopted in children's cerebral palsy and epilepsy treatment. The clinical doctor can also choose one from F1—F6 as fixed frequency.
4. Top nuclear power stimulation
4. 1 Operation time: 20 min.
4. 2 Six wave forms are selectable.
4. 3 Period: the period of wave form 1, 2, or 3 can be adjusted within 1~99s; The period of wave form 4 and 5 can be adjusted within 2~24s; The period of wave form 6 cannot be adjusted.
4. 4 Output strength: adjustable within 0~99.
5. Limbs electrical stimulation
Operation frequency: 2Hz-6Hz±10%.
Output current: adjustable within 0-100mA±5%.
The stability degree of output current is no more then ±5%.
The range of modulation frequency is not narrower than 0-150Hz.
Prescription: adjustable within 1-19.
Output strength: adjustable within 0-99.
Diathermanous temperature has 6 adjustable gears: 38°C-50°C±3°C.