AG-IIR003A intensive care equipment infant incubator price
USD $1,000 - $5,000 /Set
Min.Order:1 Set
Zhangjiagang Geer Nursing Equipment Co., Ltd.
AG-IVP003 plastic base with 5 castors Hospital intravenous drip
AG-IVP001 adjustable stainless steel hospital furniture medical drip stand
AG-SS009A CE ISO approved Stainless steel hospital drip infusion stand
AG-SC003 Multifunction folding medical bedside curtain hospital screen
AG-SC002 CE ISO approved hospital curtain medical screen folding
infant incubator price
AG-IIR003A intensive care equipment infant incubator price
1.Microprocessor based servo controlled temperature system
2.Temperature displayed by LED
3.Alarms functions : Power failure; over temperature; temperature deviation; temperature sensor failure;fan failure;
4.>37°C temperature setting
5.A keypad lock prevent inadvertent changes the setting
6.Humidity is adjustable continuously
7.Heating power displayed on panel
8.A second thermal cut-off function for more safety
9.The infant bed inclination angle is adjustable
10.Trough for pulling and pushing ,which is convenient to add the water and wash
11.Fan with low noise and long life-span
12.Temperature deviation and over temperature could be amended in front panel directly
13.Silence/reset button for alarms
14.six operating windows,two side operating windows can be swiveled
15.Heater is electrically cut-out if air temperature inside incubator exceed 40°C
16.Cabinet with three drawers
17. Phototherapy up side unit;
Main body(including the transparent hood, control system, infant bed, water trough) |
I. V. Pole |
Mattress |
Support bracket |
Castors |
Phototherapy up side unit;
Phototherapy from up and down sides;
RS232(485)conncector; Tray/pole;
Infant bed pulled from side door;
Infant bed pulled from side and front door;
Double wall hood
1.Power supply : AC 110/220V 60/50Hz
2.Power input : ≤420VA
Operating conditions:
1.Environment temperature: 20°C~30°C
2.Environment relative humidity: 30%~75%RH
3.Environment air velocity of flow: <0.3m/s
4.Air Temperature control range: 25°C~38°C
5.Temperature fluctuation: +/-0.5°C
6.Uniformity of mattress temperature: ≤0.8°C
7.Humidity accuracy: ≤10%RH
8.Warm-up time: ≤45min
9.Internal noise level: ≤55dB
10.Infant bed tilt angle : +10º
11.Mattress size: 65cm(L)×37cm(W)
12.Trough capacity: 1200ml
13.Air filter:0.5um
Environment Temp:-10°C--+70°C
Environment Relative Humidity: ≤80%
Atmospheric Pressure: 500~1060Hpa