Iec60068-2-75 Standard Laboratory Equipment Spring Hammer Test
Negotiable /Piece
Min.Order:1 Piece
Shenzhen Chuangxin Instruments Co., Ltd.
Required by many Standards to test the mechanical integrity of product enclosures. After applying the impact with the hammer, the product is examined with accessibility probes to determine access to shock, energy, and injury hazards. Built in exact accordance to IEC, EN, UL, CSA and other international Standards.
Application:
Adjust the impact severity by rotating the knurled cap until the desired energy value is displayed. Pull the knob slowly out of the device until it locks in place. If moved too fast, the clamp may skip its arrester and will lock at its way back when the pulling force declines. Hold the device horizontally at it’s housing tube and press the (black) release cone to the specimens plane of impact. The strike is automatically released when the release force threshold is reached
Types:
IK01 | IK02 | IK03 | IK04 | IK05 | IK06 | IK07 | IK08 | IK09 | IK10 |
0.14J | 0.2J | 0.35J | 0.5J | 0.7J | 1J | 2J | 5J | 10J | 20J |
Technical Parameters:
Length | 211mm |
Weight | 1250g |
OD | 50mm |
Hammer head Weight | 60g |
Hammer radius | 10mm |
The release force of impact | <10N |
Single spring impact hammer | 0.14J,0.20J,0.35J,0.50J,0.70J,1.0J 2.0J Selection |
Universal spring impact hammer(adjustable imapct hammer) | 0.14, 0.20,0.35,0.50,0.70 and 1J |
Product Details
Body: Includes release mechanism and striking element guide, mass 1250 g.