High Standard Medical Equipment Infant Radiant Warmer
USD $1,000 - $2,000 /Set
Min.Order:5 Sets
Shanghai Sun-Shore Medical Instruments Co., Ltd.
Specifications:
* Power: AC110/220V 50Hz--60Hz
* Input power: 1000VA
* Operation environment requirements:
a. Environment temperature: 18 centidegree~30 centidegree
b. Environment relative humidity: 30% RH~75% RH
c. Atmospheric pressure: 700 hPa~1060 hPa
d. Air velocity of flow:<0.3m/s
* The control range of temp: 32 centidegree~38 centidegree
* Skin temperature sensor accuracy: ±0.3 centidegree
* Temperature control accuarcy: ≤0.5 centidegree
* Uniformity of mattress temperature: ≤2.0 centidegree
* Warm-up time: ≤45min
Features:
* Servo-controlled by microcomputer
* Temperature displayed by LED
* Multiply self-checking alarms with audible and visible:
Power failure; Over temperature; Temperature deviation; Temperature sensor failure;
*Skin temperature sensor failure protect function avoid over temperature
* A second thermal cut-out function for more safety
* Heating power rate indicate 0-100%
* A keypad lock prevent inadvertent changes the setting
* Adopts anti-explosion micro crystal quartz infrared radiation tube for its radiation source, the heating is quick and
life-span is long
*APGAR timer to record therapy time
* Heater head can be adjusted +90 degree
* Tilt angle of infant bed is adjustable
* Convenient X-ray cassette under infant bed
*Observe light angle is adjustable
* LED phototherapy unit, low pressure aspirator and oxygen supply system
Optional parts:
* Drawer
* RS232(485) connector;