Basic introduction
1, accord with IEC60695-2-10, UL 746A, GB/T5169.10, GB4706.1
2, using electricity heating regulation material (Ni80 / Cr20) of resistanse wire (Φ 4 mm Ni-Cr alloy wire) to 550 °C~ 960 °C in 1 min,then with 1.0 N vertically burn it 30s,measure the fire risk by product light-off or holde burning time ,also test light-off degree, light-off temperature (GWIT), flammability and combustible index (GWFI).
3, suitable for lighting equipment, low voltage electric appliance and electrician instruments, such as electrical and electronic products..
Technical parameters
1, Glow wire temperature: 500 ~ 1000 °C± 2 °C continuous adjustable
2, Searing time: 0-999.9 s ± 0.1 s adjustable (usually selected for 30s)
3 Light-off time: 0-999.9 s ± 0.1 s, automatic recording, manual timeout
4, Fireouttime: 0-999.9 s ± 0.1 s, automatic recording, manual timeout
5, Thermocouple Φ 0.5 mm: import armoured thermocouple type (K)
6, Glow wire: Φ 4 mm Ni-Cr alloy wire
7 Ironing into depth: 7mm ± 0.5 mm
8, Sample pressure: 1N ± 0.2 N
9, Test speed: 10mm/s-25mm/s
10, Test background: black background
11, Test process: automatic control, independent convulsions
12, Consult standard GB/T5169.10-2006 , IEC60695-2-10:2000,UL746A
13, Studio volume: 0.5 cubic, stainless steel box (according to customer requirements 0.75 cubic or 1 cubic)
14, Size: 1120 x 520 x 1250mm (W*D*H)