Anaesthetic Flowmeters with Hypoxia
Negotiable
Min.Order:1
C M FLOWMETERS (INDIA) PRIVATE LIMITED
PART NAME: FLOWMETER ASSY WITH HYPOXIC GUARD (5-TUBE) | ||
1 | Oxygen Inlet | 7/16 X 20T WF male thread, from rear bottom |
2 | N2O Inlet | 1/2 X 20T WF male thread, from rear bottom |
3 | Air Inlet | 1/4 BSP male thread, from rear bottom |
4 | Mixed gas outlet | At top right, through 23mm taper as per ISO 5356-1 |
5 | Control Knob Oxygen | White, with special profile as per ISO 5358 |
6 | Control Knob N2O | Blue serrated, serration depth <1mm, knob dia lower than Oxygen knob dia. |
7 | Control Knob Air | Black serrated, serration depth <1mm, knob dia lower than Oxygen knob dia. |
8 | Working pressure (max) | 420 kPa (N2O, O2 &AIR) |
9 | Oxygen Tube Range Low Flow | 0.1 1 l/min, with Graduations in lpm unit |
10 | Oxygen Tube Range High Flow | 1 10 l/min, with Graduations in lpm unit |
11 | N2O Tube Range Low Flow | 0.1 1 l/min, with Graduations in lpm unit |
12 | N2O Tube Range High Flow | 1 12 l/min, with Graduations in lpm unit |
13 | Air Tube Range | 0.1 15 l/min, with expanded scale at 0.1 1 l/min Graduations in lpm unit |
14 | Float Colour | Red for O2, N2O & AIR |
15 | Accuracy (N2O, O2&Air) | As per CM Standard |
16 | Tube length (N2O, O2&Air) | 230 mm |
17 | Markings/identification | Float mark on tubes & other markings as per ISO 5358 |
18 | Mounting | Back bar mounting, as per drawing |
19 | Housing | Aluminum frame with clear acrylic front & Light coloured Luminous background. |
20 | N2O flow opening/closing | Opening at 0.4-0.6 l/min Oxygen flow, while increasing Closing at 0.2-0.3 l/min Oxygen flow, while decreasing |
21 | Oxygen / N2O ratio | 30 + 5%of O2 at all settings after N2O flow opening. |