MicroPette Plus Autoclavable Pipette (Adjustable & Fixed)
Negotiable /Piece
Min.Order:1 Piece
Hangzhou Chincan Trading Co., Ltd.
Summary
The MicroPette plus pipettes can be fully autoclaved and withstood steam sterilizes at 121℃, 1 atm for 20 minutes. The pipettes can be autoclaved without special preparations. After autoclaving the pipette must be cooled down and left to dry over 12 hours.
Features for MicroPette plus
- Fully autoclavable
- Lightweight, ergonomic, low force design
- Digital display clearly reads volume setting
- The pipettes cover volume range of 0.1μl to 10ml
- Easy to calibrate and maintain with tool supplied
- Design helps avoid repetitive strain injuries
- Manufactured from innovative material
- Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
Features for multi-channel MicroPette plus
- 8 and 12 channel pipettes are available for standard 96-well plate
- Dispensing head rotates for optimum pipetting convenience
- Individual piston and tip cone assemblies allow easy repair and maintenance
- Compound material tip cone design allows visual seal verification
- Can be used with universal style pipette tips
Specifications
Single-channel Adjustable Volume Pipettes | |||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||
% | μL | % | μL | ||||||
0.1-2.5μL | 0.05μL | 2.5μL | 2.50% | 0.0625 | 2.00% | 0.05 | |||
1.25μL | 3.00% | 0.0375 | 3.00% | 0.0375 | |||||
0.25μL | 12.00% | 0.03 | 6.00% | 0.015 | |||||
0.5-10μL | 0.1μL | 10μL | 1.00% | 0.1 | 0.80% | 0.08 | |||
5μL | 1.50% | 0.075 | 1.50% | 0.075 | |||||
1μL | 2.50% | 0.025 | 1.50% | 0.015 | |||||
2-20μL | 0.5μL | 20μL | 0.90% | 0.18 | 0.40% | 0.08 | |||
10μL | 1.20% | 0.12 | 1.00% | 0.1 | |||||
2μL | 3.00% | 0.06 | 2.00% | 0.04 | |||||
5-50μL | 0.5μL | 50μL | 0.60% | 0.3 | 0.30% | 0.15 | |||
25μL | 0.90% | 0.225 | 0.60% | 0.15 | |||||
5μL | 2.00% | 0.1 | 2.00% | 0.1 | |||||
10-100μL | 1μL | 100μL | 0.80% | 0.8 | 0.15% | 0.15 | |||
50μL | 1.00% | 0.5 | 0.40% | 0.2 | |||||
10μL | 3.00% | 0.3 | 1.50% | 0.15 | |||||
20-200μL | 1μL | 200μL | 0.60% | 1.2 | 0.15% | 0.3 | |||
100μL | 0.80% | 0.8 | 0.30% | 0.3 | |||||
20μL | 3.00% | 0.6 | 1.00% | 0.2 | |||||
50-200μL | 1μL | 200μL | 0.60% | 1.2 | 0.15% | 0.3 | |||
100μL | 0.80% | 0.8 | 0.30% | 0.3 | |||||
50μL | 1.00% | 0.5 | 0.40% | 0.2 | |||||
100-1000μL | 5μL | 1000μL | 0.60% | 6 | 0.20% | 2 | |||
500μL | 0.70% | 3.5 | 0.25% | 1.25 | |||||
100μL | 2.00% | 2 | 0.70% | 0.7 | |||||
200-1000μL | 5μL | 1000μL | 0.60% | 6 | 0.20% | 2 | |||
500μL | 0.70% | 3.5 | 0.25% | 1.25 | |||||
200μL | 0.90% | 1.8 | 0.30% | 0.6 | |||||
1000-5000μL | 50μL | 5000μL | 0.50% | 25 | 0.15% | 7.5 | |||
2500μL | 0.60% | 15 | 0.30% | 7.5 | |||||
1000μL | 0.70% | 7 | 0.30% | 3 | |||||
2-10mL | 0.1mL | 10mL | 0.60% | 60 | 0.20% | 20 | |||
5mL | 1.20% | 60 | 0.30% | 15 | |||||
2mL | 3.00% | 60 | 0.60% | 12 | |||||
8-channel Adjustable Volume Pipettes | |||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||
% | μL | % | μL | ||||||
0.5-10μL | 0.1μL | 10μL | 1.50% | 0.15 | 1.50% | 0.15 | |||
5μL | 2.50% | 0.125 | 2.50% | 0.125 | |||||
1μL | 4.00% | 0.04 | 4.00% | 0.04 | |||||
5-50μL | 0.5μL | 50μL | 1.00% | 0.5 | 0.50% | 0.25 | |||
25μL | 1.50% | 0.375 | 1.00% | 0.25 | |||||
5μL | 3.00% | 0.15 | 2.00% | 0.1 | |||||
50-300μL | 5μL | 300μL | 0.70% | 2.1 | 0.25% | 0.75 | |||
150μL | 1.00% | 1.5 | 0.50% | 0.75 | |||||
50μL | 1.50% | 0.75 | 0.80% | 0.4 | |||||
12-channel Adjustable Volume Pipettes | |||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||
% | μL | % | |||||||
0.5-10μL | 0.1μL | 10μL | 1.50% | 0.15 | 1.50% | ||||
5μL | 2.50% | 0.125 | 2.50% | ||||||
1μL | 4.00% | 0.04 | 4.00% | ||||||
5-50μL | 0.5μL | 50μL | 1.00% | 0.5 | 0.50% | ||||
25μL | 1.50% | 0.375 | 1.00% | ||||||
5μL | 3.00% | 0.15 | 2.00% | ||||||
50-300μL | 5μL | 300μL | 0.70% | 2.1 | 0.25% | ||||
150μL | 1.00% | 1.5 | 0.50% | ||||||
50μL | 1.50% | 0.75 | 0.80% | ||||||
Fixed Volume Pipettes | |||||||||
Volume Range | Increment | Test Volume | Maximum permissible systematic error (Inaccuracy) | Maximum permissible random error (Imprecision) | |||||
% | μL | % | |||||||
5μL | - | 5μL | 1.30% | 0.065 | 1.20% | ||||
10μL | - | 10μL | 0.80% | 0.08 | 0.80% | ||||
20μL | - | 20μL | 0.60% | 0.12 | 0.50% | ||||
25μL | - | 25μL | 0.50% | 0.125 | 0.30% | ||||
50μL | - | 50μL | 0.50% | 0.25 | 0.30% | ||||
100μL | - | 100μL | 0.50% | 0.5 | 0.30% | ||||
200μL | - | 200μL | 0.40% | 0.8 | 0.20% | ||||
250μL | -- | 250μL | 0.40% | 1 | 0.20% | ||||
500μL | - | 500μL | 0.30% | 1.5 | 0.20% | ||||
1000μL | - | 1000μL | 0.30% | 3 | 0.20% | ||||
2000μL | - | 2000μL | 0.30% | 6 | 0.15% |
Normally all CHINCAN products are supplied with a warranty of 12 Months from the date of shipment. This warranty is entirely dependent on the following conditions:
1.The following circumstance don't belong to the warranty policy which is Man-made damages caused by improper operations.
2.The instruments must be operated in accordance with our operating guidelines; serviced and maintained on a regular basis in accordance with the terms specified in the relevant user manual.
3. Only approved spare parts should be used in the instruments which bought from CHINCAN and these should be changed on a regular basis as specified in the relevant user manual. Details of the approved spare parts and accessories for each instrument are shown in the user manual or alternatively this information can be obtained from CHINCAN.
Q: What after-sales service can be provided by CHINCAN ?
A: The after-sales service of CHINCAN is divided into two parts: technical support and maintenance.
Q: What about the technical support and maintenence?
A: Sometimes users will get some problems when they operate the instruments which are unfamiliar or first time use, users can send us emails or call us about detailed information. We will provide users solution or operation video.For so many years’ experience, our instruments works well during warranty period. If there any parts are broken during warranty period, we would like to provide spare parts for free.
Q: How to solve the disputes between CHINCAN and customers during the after-sales maintenance?
A: We will be customer oriented,and make concessions in some principles , we will properly handle the disputes to maximize customer's satisfaction.
Established in 2004, Hangzhou Chincan Trading Co., Ltd. is a foreign trading company authorized by Chinese State Department of Foreign Economy & Trade and General Administration of Chinese Customs, which located in Hangzhou city, Zhejiang province, China.
We are a specialist company committed to exporting laboratory instruments, educational apparatus, medical & hospital equipments. With its superior performance, Chincan instruments are sold well throughout the Americas, Asia Pacific, the Middle East, Africa and Europe. With our excellent service, Chincan Trading has won a good reputation from the market.
Cooperative factories which qualified by Chincan come from many provinces and cities across China. All of them need to be ISO certified, Some products have CE.