Product Details

Jumper hot sale portable multi parameter patinet monitor with ETCO2 option

Type First-Aid Devices
Brand Name Jumper
Model Number JPD-800B-a
Place of Origin Guangdong, China

Product Features

12inch Veterinary monitor--suitable for many kinds of animals

Features:

 12" color TFT display, up to 12 waveforms
One level menu, easy to use
Parameters including ECG/RESP, SpO2 , NIBP, TEMP, Pulse Rate and IBP(optional)
Maximum 720 hour graphic and tabular trends of all parameters
Built-in recorder and battery
Suitable for dog,cat, horse ,bird etc

Arrhythmia analysis, ST calculation, Pace analysis
Anti-ESU, anti- defibrillator
Standard network interface, can connect to Central monitor system
Optinal: recorder, dual IBP, CO2, IoC

 

 Jumper hot sale portable multi parameter patinet monitor with ETCO2 optionJumper hot sale portable multi parameter patinet monitor with ETCO2 optionJumper hot sale portable multi parameter patinet monitor with ETCO2 option

Specifications

 

 ECG

Lead mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V)
Lead selection I, II, III, avR, avL, avF, V,
Waveform 2 ch
Lead mode 3 Leads ( R, L, F or RA, LA, LL)
Lead selection I, II, III,
Waveform 1 ch
Gain: X2.5mm/mV, X5.0mm/mV, X10mm/mV, X20mm/mV, auto
Resolution 1 bpm
Sensitivity > 200 (uV P-P)
Differential Input Impedance > 5 MΩ
CMRR
Monitor > 105 dB
Operation > 105 dB
Diagnosis > 85 dB
Electrode offset potential 300mV
Leakage Current < 10 uA
Baseline Recovery < 3 S After Defi.
ECG Signal Range: 8 m V ( Vp-p )
Bandwidth
Surgery 1 ~ 15 Hz
Monitor 0.5 ~ 35 Hz
Diagnostic 0.05 ~ 100 Hz
5%±Calibration Signal 1 (mV p-p), Accuracy :
ST Segment Monitoring Range
Measure and Alarm -2.0 ~ +2.0 mV
ARR Detecting

 

Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R
ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Alarm Available
Review Available


RESPARATION
Method Impedance between R-F(RA-LL)
Differential Input Impedance >2.5 MΩ
Measuring Impedance Range: 0.3~5.0Ω
Base line Impedance Range: 0 ~ 2.5 KΩ
Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate
Measuring and Alarm Range Adult 0 ~ 120 rpm
Neo/Ped 0 ~ 150 rpm
Resolution 1 rpm
Accuracy  2 rpm
Apean Alarm 10 ~ 40 S


NIBP
Method Oscillometric
Mode Manual, Auto, STAT
Measuring Interval in AUTO Mode
1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
Measuring Period in STATMode 5 Min
Pulse Rate Range 40 ~ 240 bpm
Alarm Type SYS, DIA, MEAN
Measuring and alarm range
Adult Mode
SYS 40 ~ 270 mmHg
DIA 10 ~ 215 mmHg

MEAN 20 ~ 235 mmHg
Pediatric Mode
SYS 40 ~ 200 mmHg
DIA 10 ~ 150 mmHg
MEAN 20 ~ 165 mmHg
Neonatal Mode
SYS 40 ~ 135 mmHg
DIA 10 ~ 100 mmHg
MEAN 20 ~ 110 mmHg
Resolution
Pressure 1mmHg

  

Accuracy
Pressure
Maximum Mean error  5mmHg
Maximum Standard deviation  8mmHg
Overpressure Protection
Adult Mode 297 3 mmHg
Pediatric Mode 240 3 mmHg
Neonatal Mode 147 3 mmHg

 

SpO2
Measuring Range 0 ~ 100 %
Alarm Range 0 ~ 100 %
Resolution 1 %
Accuracy
70% ~ 100%  2 %
0% ~ 69% unspecified
Actualization interval about 1 Sec.
Alarm Delay 10 Sec.
Pulse Rate
Measuring and Alarm Range
0~254bpm
Resolution 1bpm
Accuracy    2bpm


TEMPERATURE
Channel 1
Measuring and Alarm Range 0 ~ 50 C
Resolution 0.1 C
Accuracy  0.1 C
Actualization interval about 1 Sec.
Average Time Constant < 10 Sec.

 
IBP(Optional)
Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measuring and alarm range
ART 0 ~ 300 mmHg
PA -6 ~ 120 mmHg
CVP/RAP/LAP/ICP -10 ~ 40 mmHg
P1/P2 -10 ~ 300 mmHg

Press Sensor
Sensitivity 5 uV/V/mmHg
Impedance 300-3000Ω

 

 

Resolution 1 mmHg
Accuracy 2% or 1mmHg, which great
Actualization interval about 1 Sec.

  

ETCO2(Optional)
Measure range: 0% - 13%
Resolution: 1 mmHg
Accuracy : ±2 mmHg (< 5.0% Measurement Value)
±10%( > 5% Measurement Value)
Response Time: 180ms

 

  