Specifications
1.Gas Transmisson Rate Analyzer
2. Gas(N2,CO2,H2,He,etc.) transmission rate (GTR) of flexible materials test
Gas Transmisson Rate Analyzer,GTR testing machine
Technical Specifications
.
Measurement range | 0.02~10000 g/m2.day |
Measurement accuracy | 0.001 g/m2.day |
Temp. range | 10°C~50°C |
Temp. accuracy | 0.1°C |
Humidity range | 0~20%RH seasoning,10~90%RH(dual-airflow humidity method, selection) |
Humidity accuracy | ±2%RH |
Sample thickness | ≤2mm |
Sample number | 3pieces |
Sample dia | Φ80mm ,transmission area 50.24cm2,(no stipulate for area in international) |
Features
- N 530 has three test chambers with three separate SENSORS and are self standing instruments. It gives simultaneous test reports, as all the test chambers have separate SENSOR .
- N-Series GTR adopts the international advanced Electro- magnetic technology , using the processing step controlling system to set the temperature. Within the range of 5~95,both raising and lowering the temperature can be done. Temperature coule be set to the exact value, and the velocity could be adjusted. A security mode is designed inside so as not to lose control.
- Exact temperature controlling method.The temperature controlling device and sensor are directly embedded on the chamber, so as to ensure the temperature controlling precision, prevent the temperature transmitting delay.
- Adopts double flow method to control the humidity.This method is an international advanced humidity control method which features steady humidity and high precision. (optional)