2016 Anesthesia Machine with Ventilator (1 Vaporizer,2 Gas)
Negotiable /Unit
Min.Order:1 Unit
Anjue Medical Equipment Co., Ltd.
Anesthesia Machine with Ventilator AJ-2101A
AJ-2101A is an economic type of Anesthesia Machine.
The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose one kind of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.
Reliable pneumatic transmission system.
Economic and proper configuration of technologies, optional as required by the Customer.
Modularized design and upgrading. Upgrading is available as required by the Customer.
TECHNICAL PARAMETERS
Part of Anesthesia
*Main body High-strength engineering plastic rack, light,
beautiful and corrosion resistant
*Scope of application adult, child
*Gas source O2, N2O: 0.27~0.55MPa
*Flow meter O2: 0.1~1.0L/min 1.1~10L/min
*Flow rate of rapid oxygen supply 25~75L/min
*Low oxygen pressure alarm There will be sound alarm when the oxygen
pressure < 0.2MPa
*Vaporizer It has the function of automatic compensation based on pressure,temperature,and flow rate. The regulation range of evaporator concentration is 0~5 vol%.
Among Halothane,Enflurane,Isoflurane and Sevoflurane,one can be chosen for application as required by the customer.
*Respiratory circuit working mode: all-close, semi-close, semi-open
releasingpressure: 0.5~7 kPa
*Respiratory bellows bellows for adults, bellows for children
tidal volume range:0~1500 ml
Part of Ventilator
*Display mode High-definition 4.8” LCD screen display
*Ventilation mode IPPV, MANUAL (Manual monitoring of tidal volume, ventilation volume, respiratory rate)
*Ventilation parameters Tidal volume 50~1500ml
Rate 4~80 bpm
I:E 2:1~1:4
Pressure range 1.0~6.0 kPa
*Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, I/E, peak pressure of airway,inspiratory trigger pressure, pressure - time waveform
*Security Alarm System
Airway pressure alarm upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa
TV alarm upperlimit setting range 0.5~2.0L
low limit setting range 0.02~1.0L
Sustained high-pressure alarm it will give alarm when stress have consistently
been higher than 2.5 kPa
Suffocation alarm it will give sound and light alarm if there is
no tidal volume input in 15s
Power alarm
*Power AC 220V±10% 50Hz±1Hz