A/B scanner NDM2100
Negotiable
Min.Order:1
Wenzhou Jinzhou Group Foreign Trade Industries Co., Ltd.
B MODE:
Frequency: 10MHZ
Gain control: ≥98dB
TGC: -30dB-0dB
Resolution: axial≤0.2mm; lateral≤0.4mm
Scanning scope:
angle: 53°
depth: from 34mm to 60mm
Delay control: ≥0mm-15mm
Gray scale: 256
Cineloop: 5.6 sec/ 56 images; single or circle
Memory: 8 images
A MODE :
Frequency: 10MHz
Measuring mode: immersion/contact
Biometry accuracy: ≤0.06mm
Display resolution: 0.01mm
Measuring scope(AL): 16—40mm
Measuring parameter: corneal thickness (immersion mode), anterior chamber depth,
Lens thickness, vitreous length and axial length
Measuring method: automatic ( for normal, aphakic, special, dense cataract) and manual; special for any kinds of pseudolens and silicon oil eye, etc
IOL calculation: SRK/T, SRK- Ⅱ ,BINKHOST- Ⅱ ,HOLLADAY, HOFFER-Q and HAIGIS
Memory: up to 50 groups data ,permanently
Others:
Image Output (option): Video printer or Workstation
Power supply: AC 100V-240V, 50/60 Hz
Dimensions: 360×300×260mm
Weight: 10kg
FEATURES:
higher resolution
B,B+B,B+A,A features
Zoom from 34mm—60mm
Variable displaying area
Adjustable TGC
Cineloop for 5.6sec
8 images memory
Image processing
Multi-groups of calipers for distance measurement
Vary area measurement
With immersion mode measurement
Save for up to 50 groups A measurement results
Trackball for easy measurement and operation
Full screen character note