Technique item | Technique specifications: |
Overall maximum size | 280*130*120mm (L*W*H) |
Gross weight | <3.0kg |
Classification | Class II, type BF |
Power supply: | AC100-240V, 50/60HZ |
Fuse | 2 pcs of 220V0.5A, 2 pcs of 1.5A |
Max power consumption | 25VA |
Applicable syringes | Automatic identification of 5ml, 10ml, 20ml, 30ml and 50ml. |
Injection rate range | 5ml: 0.1ml/h---100ml/h; 10ml: 0.1ml/h---200ml/h |
20ml: 0.1ml/h---400ml/h; 30ml: 0.1ml/h---600ml/h |
50ml: 0.1ml/h---1500ml/h |
Rate increment | 0.1ml/h |
Injection accuracy: | ±3%( after being adjusted, up to ±2%) |
KVO rate | Adjustable range: 0.1ml/h-5.0ml/h, choose “OFF” to cancel KVO function if no need KVO (entering into KVO after OCCUL when running) |
Bolus rate | 5ml:100ml/h 10ml:200ml/h 20ml:400ml/h 30ml:400-600ml/h 50ml:400-1500ml/h |
Selectable modes | 1. Rate mode 2. Time mode 3. Body weight mode |
Rate mode | Rate: 0.1-1500ml/h (max value depends on the syringe size) |
Time mode | Volume limit: 0.1-999.9ml time: 1-2000min |
Body weight mode | Body weight: 0.1~300kg drug limit: 0.1~999.9mg volume limit: 0.1~999.9ml Dose: 0.1~9999.9 (subject to unit, dosage, liquid, weight and syringe) unit: mg/kg/h, ug/kg/min |
Accumulated volume | 0.1~9999.9ml |
Preset range | 0.1~9999.9ml |
Preset time | 0-99h59min |
Drug library | More than 20 drug names: Penicillin, Vancomycin, Pitocin, Insulin, Lidocaine, Heparin, Atropin, Dopamine, Epinephrine etc. |
Alarm Function | NEAR, OVER, OCCL., syringe disengaged, no AC power supply, low battery, serious low battery, error 1/2/3, preset volume alarm |
OCCUL value | 0.04Mpa-0.16Mpa |
Battery | Rechargeable lithium polymer battery 7.4V, 1600mA Working for 4 hours (5ml/h) after fully being charged |