Product Details

IEC61000-4-11 CSS-20P1 Voltage Dips and Short Interruption Generator

Place of Origin China
Model Number CSS-20P1
Power Electronic

Product Features

IEC61000-4-11 CSS-20P1 Voltage Dips and Short Interruption Generator

1. Introduction:

Voltage dips and short interruption generator is designed according to IEC6100-4-11, it is to simulate  effects of abrupt voltage change, includes voltage dips and short interruption. The voltage dips and short interruption immunity tests are applied to electrical and electronic equipment having a rated input current not exceeding 16A per phase, and for connection to 50Hz or 60Hz AC networks. This is a basic EMC test to assess the immunity of equipment or a system against the voltage dips and short interruptions caused by fault in the network, primarily short circuits, in installation or by sudden large changes of load.

 

2. Features:

 Using 2 voltage regulators, the rated voltage can be adjusted and automatically stabilized, which can avoid the influence of power grid fluctuations on the test results. Simulate the test under 110V power supply. Can simulate the situation of power grid fluctuations under extreme conditions (such as the power supply voltage in remote areas is relatively high or low)

 Adopts 7 inch LCD touch screen, Waveform graphic display, easy to operation.

 EUT output comes with high-strength panel terminals, which makes the connection of test samples more convenient and reliable

 The instrument panel has external voltage and ammeter output, which is convenient to check the working status of the instrument.


3. Parameters:

Model

CSS-20P1

CSS-20P3

Control and display mode

LCD+touch screen

Load capacity

Single phase 220V/20A ( 2 regulators)

3 phase 380V/20A (3 regulators)

Voltage dips

0~265V±10% (1V step), the upper limit is limited by the grid voltage

0~265V±10% (1V step), each phase is set independently. The dip voltage can be different to simulate three-phase unbalance.

Rated voltage

90~265V±10% (1V step), the upper limit is limited by the grid voltage


Rise (and fall) initial phase

0~359°(1°step)

Rise (and fall) end phase

0~359°(1°step)

IEC level built in:

Voltage rise (and fall) 0%, 40%, 70%, 80%, 120% of UT

Trigger mode:

COUNT

Count according to the set value of the falling (rise) cycle

50ms

Each cycle generated by the simulator includes three rises (and falls), and the time interval between these three rises (and falls) is 50ms

Duration for voltage dip

1~9999ms

1~9999ms, each phase can be set independently

Interval for voltage dip

1~9999ms

1~9999ms, each phase can be set independently

Rise (and fall) cycles

1~9999 times

Working condition

Temperature:10℃~35 Humidity: 30%~60%

Power:

Single phase, AC220V±10%,50/60Hz

Weight:

Approximate 100kg

Approximate 200kg