High Temperature Burn in Accelerated Aging Test Room
Uses:
Walk-in constant Temperature and humidity laboratory for industry large parts, semi-finish product,finish products
temperature and humidity environment testing. Such as computer terminal, car nstant temperature
and humidity room is composed of control panel,switchboard,hydrating panel,fan, heater,humidifier,refrigerator
This equipment mainly used for create and keep temperature humidity constant space,completely automatic control
refrigeration,heating,dehumidify,humidification,supplementary humidified water,The equipment can be added (optional)
air exchange function,operator can long term work in it when exist poor volatile gas.Suitable for electrical, electronic,
instrument and other products, parts and components salt spray test chamber and material compatibility test of storage
transport using under high low temperature alternating,humid and hot environment.
Design Standard:
IEC60068-2-1.1990 testing methods for low-temperature
IEC60068-2-2.1974 testing methods for high-temperature
GJB150.
Character:
- Composed of control panel,switchboard,hydrating panel,blower, heater,humidifier,refrigerator, mainly specification divided into 7 standards indoor dimension and 28 different conditions testing standards
- Adopt panel unit combination. Inner Volume can be arbitrarily amplify,easy dismounting,cooperate and as clients request dimension and design relocation of expanse plant.
- Board use stainless steel and Meinai steel plate with baking paint,firm structure,water proof, beauty appearance.
- Adopt multi-blade centrifugal air supply mode horizontal spread vertical heat exchange circulation system,avoid any blind spot,make room temperature humidity uniform distribution.
- Abnormal condition occurs,screen will automatic display failure state,turn off power and provide trouble shooting method.
- Refrigerator use western original import compressor, and use environmental friendly refrigerant,refrigeration system use binary type low temperature loop design and different temperature adopt different power of compressor to save energy and service life.
Image:
Parameter:
High Temperature Burn in Accelerated Aging Test Room
Model |
RHPW-4 |
RHPW-6 | RHPW-8 | RHPW-13 | RHPW-17 | RHPW-25 |
System | Balanced temperatuer & humidity control system | |||||
Testing temperature range |
R.T-65℃~+120℃ |
|||||
Humidity range |
20%-95%RH |
|||||
Temperature fluctuation |
±0.5℃ |
|||||
Humidity fluctuation |
±2.5%RH |
|||||
Temperture uniformity | ±2.0℃ | |||||
Humidity uniformity |
±5.%RH |
|||||
Temperture deviation |
±2℃ |
|||||
Humidity deviation |
±3%RH (≥75%) ±5%RH(<75%) |
|||||
Heat up time |
20℃~+80℃ within 60 minutes | |||||
Pull down time |
20℃~-55℃,-40℃,-30℃,-20℃,-10℃ within 120 or 90 or 60 minutes. | |||||
Interior volume(m3) | 9.04 | 12.9 | 16.78 | 27.09 | 35.21 | 43.34 |
Interior material | SUS#304 Stainless steel plate | |||||
Exterior material | Model steel | |||||
Insulation material | Rigid polyerethane foam +glass fibre | |||||
Floor bearing load | 500Kg/m3 | |||||
Refrigeration system | Air cooled (Water cooled type is option) | |||||
Cascade type, hermetic/semitight compressor,CFC free refrigerant | ||||||
Controller | Color LCD screen in English/Chinese, programing,communicating function | |||||
Safety devices | No welding wire switch,compressor overheating and over current protection,over temperature protection, blower overload protection,dry heating machine protector,low water protection,fault warning system. | |||||
Power | AC380±10% 50HZ 3 phase 4 ires+ground wires |
Optional accessories:
- Test hole (Cable prot)
- Double open gate (Doors)
- Buffer room (Pass room)
- Mobile slope platform (Carry-in inclined)
- Temperature and humidity recorder (Reolving lamp)
- Ceiling blowing type (Entire ceiling blow-out duct)
- State indicator
- Pure water unit (Water purifier)
- Power outlet (Plug socket)
- Low humidity dehumidification equipment (Dehumidifier)