Product Details

PID-A12 Photo Ionisation Detector

Product Features

PID-A12 Photo Ionisation Detector

 

PERFORMANCE
Target gases          VOCs with ionisation potentials                 < 10.6 eV
Minimum resolution      ppb isobutylene                                                < 50
Linear range          ppm isobutylene 5% deviation                           300
Overrange            ppm isobutylene                                               6000
Sensitivity linear range  mV/ppm Isobutylene, see Table 1 for options   > 0.3
Full stabilisation time   minutes to 100 ppb                                               20
Warm up time          seconds time to full operation                                 5
Offset voltage         mV                                                               50 to 59
Response time (t90)      seconds diffusion mode                                       < 3

 

ELECTRICAL
Power consumption   85 mW (max) at 3.2 V, 350 mW transient for 200 msec on switch-on
              90 mW at 3.3 V, 460 mW transient for 200 msec on switch-on
Supply voltage     3.2 to 3.6 VDC Ideally regulated ±0.01V (onboard regulator disabled)
              3.6 to 10 VDC (onboard regulator enabled)
              (maximum 10V for IS approval, maximum 18 V for non-IS)
Output signal      Offset voltage (minimum 50 mV) to Vmax
              (Vmax = Vsupply -0.1 V when regulator is enabled)

 

ENVIRONMENTAL
Temperature range    -40℃ to +55℃(Intrinsically Safe); -40℃ to +65℃(non-IS)
Temperature dependence 0℃ to 40℃ 90% to 100% of signal at 20℃
               -20℃ 140% of signal at 20°℃ 
Relative humidity range Non-condensing 0 to 95%
Humidity sensitivity   During operations: 0% to 75% rh transient near zero

 

KEY SPECIFICATIONS
Operating life    5 years (excluding replaceable lamp and electrode stack)
IS Approval      IECEx Ex ia IIC T4;ATEX Ex ia II 1G -40℃ < Ta < +55℃ (< 10VDC supply)
Onboard filter    To remove liquids and particulates
Lamp           User replaceable
Electrode stack    User replaceable
Error state signal  Lamp out: 32 ±4 mV
              Electronic error: 22 ±6 mV
Weight           < 8g
Position sensitivity None 
Warranty period    Electronics and housing: 24 months
              Lamp and electrode stack are user replaceable. 10.6eV lamp: 5000 lit hours



You May Like

Find Similar Products By Category

You May Like

View More
Chat Now Contact Now