Multi-parameter water quality meter can measure simultaneously pH/pX, Ion concentration, ion electrode potential (mV) value, Condutive (conductivity, TDS, salinity), DO (concentration and saturation) and temperature, etc.
We Can Supply All Kinds Of Water Analysis Instrument To Meet Different Demands. Inquiry us to chooes the best one for you.
pH
² 1-6 points calibration with Standard Recognition.
² Selectable pH buffer groups, including NIST, DIN, GB.
² Support for pH standard group customization.
² Automatic electrode diagnosis with pH slope and offset display.
Ion
² 1-6 points calibration
² Selectable measurement unit, including μg/L, mg/L, g/L, mmol/L, pX, ppm, etc.
² Multi-measurement modes are supported, including Direct Reading mode, Standard Addition mode, Sample Addition mode and GRAN mode.
² Over 10 methods are built-in, including F-, Cl-, Br-, I-, NO3-, BF4-, NH4+, K+, Na+, Ca2+, Cu2+, Pb2+, Ag+ and etc., user-defined method is supported.
Conductivity
² 1-5 points calibration with Standard Recognition.
² Settable parameters, including cell constant, reference temperature (5/10/15/18/20/25°C), temperature compensation coefficient and TDS factor.
² Temperature compensation type (none, linear, pure water).
DO
² Support for air-saturated water or zero oxygen calibration.
² Auto barometric pressure compensation.
² Manual Salinity Factor Correction.
² Selectable pressure unit, including kPa, mbar, Torr, Atm.
pH
|
Range |
-2.000 to 20.000 pH |
Resolution |
0.1, 0.01, 0.001pH |
|
Accuracy |
±0.002 pH |
|
Calibration Points |
Up to 6 |
|
Standard Customization |
Yes |
|
Calibration Reminder |
Yes |
|
Standard Recognition |
NIST, GB, DIN buffers |
|
Slope Limit |
Yes |
|
mV/ORP
|
Range |
-2000.00 to 2000.00 mV |
Resolution |
0.1, 0.01 mV |
|
Accuracy |
±0.1 mV or ±0.03% |
|
EH ORP Mode |
Yes |
|
ORP Calibration Points |
1~2 custom point (relative mV) |
|
pX
|
Range |
-2.000 to 20.000 pX |
Resolution |
0.1, 0.01, 0.001 pX |
|
Accuracy |
±0.002 pX |
|
Calibration Points |
Up to 6 |
|
ISE
|
Range |
0~19990 |
Resolution |
Up to 4 significant digits |
|
Accuracy |
±0.3% |
|
Calibration Points |
Up to 6 |
|
Conductivity |
Range |
0.000 μS/cm to 3000 mS/cm |
|
Resolution |
0.001 μS/cm minimum, various with range selection |
|
Accuracy |
±0.5% FS |
|
Reference Temperature |
5, 10, 15, 18, 20, 25 ºC |
|
Calibration Points |
Up to 5 |
|
Standard Recognition |
146.5μS/cm, 1408μS/cm, 12.85mS/cm, 111.3mS/cm |
Resistivity
|
Range |
5.00 Ω·cm~100.00 MΩ·cm |
Resolution |
0.01 Ω·cm minimum |
|
Accuracy |
±0.5% FS |
|
TDS
|
Range |
0.000 mg/L~1000 g/L |
Resolution |
0.001ppm minimum, changed with range |
|
Accuracy |
±0.5%FS |
|
Salinity
|
Type |
NaCl% |
Range |
(0.00~8.00)% |
|
Resolution |
0.01% |
|
Accuracy |
±0.1% |
|
Dissolved Oxygen Concentration - Polarographic
|
Range |
0.00 to 99.99 mg/L |
Resolution |
0.01 mg/L |
|
Accuracy |
±0.1 mg/L |
|
Calibration Reminder |
Yes |
|
Calibration Points |
Air-saturated water or zero point |
|
Barometric Compensation |
Yes |
|
Manual Salinity Factor Correction |
Yes |
|
% Saturation -Polarographic
|
Range |
(0.0 to 600)% |
Resolution |
0.10% |
|
Accuracy |
±2.0% |
|
Temperature
|
Range |
-5 to 130 ºC |
Resolution |
0.1 |
|
Accuracy |
±0.1 |
Inquiry Us For More Details