Product Details

MC-LC98I Amino Acid Liquid Chromatography

Place of Origin Guangdong, China
Brand Name MeCan
Model Number MC-LC98I

Product Features

MC-LC98I Amino Acid Liquid Chromatography

 

model:MC-LC98I

 

 

 

MC-LC98I Amino Acid Liquid Chromatography

No.

          Type

Unit

Quantity

1

P98I Binary Gradient Pump

set

1

2

UV98I UV Detector

set

1

3

Rheodyne 7725i Injector

set

1

4

C18, 5μm, Amino Acid Column

set

1

5

Guard Column

set

1

6

N2000 Work Station

set

1

7

T98 Column Oven

set

1

8

Solvent Filter

set

1

9

0.45μ Organic Filter membrane

piece

1

10

0.45μ Water Filter membrane

piece

1

11

Organic Sample Filter

piece

1

12

Water Sample Filter

piece

1

13

Aspirator Vacuum Pump

set

1

14

Amino Acid Standard Sample

set

1

15

Amino Acid Derivatives Solvent

set

1

 

P98I Binary Gradient Pump

1. SPECIFICATION

   a. Operating pressure: 0~42MPa (6000 psi) maximum.

b. Flow rate range: 0.1~15ml/min;

c. Flow rate precision: ±1%;

d. Flow rate accuracy: 25ml/368~387s(±3%);

   e. Compositional precision:0.5% absolute

f. Pressure solution: 0.1MPa;

   g. Over pressure protection: pump shuts off if pressure exceeds limit specified by user-42.0 MPa(6000 psi) maximum.

2. Environment

a.Temperature requirements: ambient temperature range 10 to 35

b. Operates in relative humidity from 5% to 80%;

c. Power requirements: 220Vac±5%,

d.Power frequency: 50±0.5Hz (voltage and frequency specified on the tag attached to the power cord)

3. GENERAL PARAMETERS

   a. Power Supply: 400W;

b. Dimensions: 450mm(D)×210mm(W)×390mm(H);

c. Weight: 20kg;

 

UV98I UV Detector

1. Specification

Wavelength range:190-700nm

Spectral bandwidth:5nm

Wavelength accuracy:±2nm

Wavelength reproducibility:±0.2nm

Absorbance range:

Recorder output:0.002au/mV to 2au/mV in steps of 1, 2, 5

Data system output:2au/V, fixed

Noise:5×10-5au (Peak-Peak)

Test Conditions: Wavelength: 254nm

Mobile phase: methanol

Flow rate: 1ml/min

Time constants: 0.5s

Drift:1×10-3au/hr

Test Conditions: Wavelength: 254nm

Mobile phase: methanol

Flow rate: 1ml/min

Time constants: 0.5s

Flow cell volume:4.5μl

Time constants:0.05s, 0.5s, 5s

Autozero:Yes

2. General Parameters

   a. Power Supply: 200W;

b. Dimensions: 450mm(D)×180mm(W)×390mm(H);

c. Weight: 15kg;

 

MC-LC98I Amino Acid Liquid Chromatography

 

 

 

Figure 1 18 Amino Acid Stand Sample

1. Asp 2.Glu 3. Ser 4. Gly 5 His 6. Arg  7. Thr 8. Ala 9. Pro 11. Val

12. Met 13. (Cys)2 14. Ile 15. Leu 16. Trp 17. Phe 18. Lys

 

 


 

T98 Column Oven

1. Specification

   Temperature range: ambient ~150℃.

    Resolution: 1℃ steps

   Precision: ±0.5℃

2. General Parameters

   Power Supply: 100W;

Dimensions: 450mm(D)×260mm(W)×150mm(H);

Weight: 5kg;

 

 

 

 

 