Product Details

Biological Safety Cabinet-B2 supply

Brand Name Kenton
Place of Origin China
Model Number B2-1000G(D)

Product Features

Biological Safety Cabinet Selection

Type of protectionBSC Selection
Personnel Protection, microorganisms in Risk Groups 1-3Class I, Class II, Class III
Personnel Protection, microorganisms in Risk Group 4, Glove box laboratoryClass III
Personnel Protection, microorganisms in Risk Group 4, suit laboratoryClass I, Class II
Product protectionClass II, Class III
Volatile radionuclide/chemical protection, re-circulated back to work zoneClass II Type B1, Class II Type A2 vented to outside
Volatile radionuclide/chemical protection, re-circulated back to work zoneClass I, Class II Type B2, Class III vented to outside

CLEANFLOW™ Clean air technology

●Equipped with German EBM high air flow and low energy consumption fan, based on airflow mechanics principle, arc-shaped large-diameter inlet design to reduce turbulence generated, to ensure low noise operation of the system.
●Reasonable duct design, optimization of the plenum tank, with the design of micro-shaped mesh layer outflow wind deflector, non-convex, no dead ends, to ensure uniformity and stability of the airflow.
●Equipped with long life exhaust HEAP and down flow ULPA, >99.995% efficient at 0.12 micron sizes.
●BIOCASION series BSC have two types:
A2 Type:70% inner circulation, 30% exhaust.
B2 Type:100% exhaust.
Work zone width:1000mm and 1300mm

SAFELOGIC™ Plenum leak-proof technology

●Around the work zone at the top and side wall plenum design, so that the discharge of polluting gases completely in a plenum environment, work with indoor backplane to a molding design, effective way to avoid the possibility of leakage.
●Equipped with the accurate pressure sensor to monitor the pressure of the working area, where there is abnormal situation such as breakage of the filter zone of negative pressure seal failure, the system will automatically stop the fan running, and through sound and light alarm prompts the operator, thus avoiding due to leakage caused damage.

BIOCASION™ Microcomputer controller system

●The integrated LCD screen. Display inflow/down flow speed, pressure, filter life, failure alarm and other parameter status.
●The user-friendly design, with sterilization, lighting, interlock function, frond window over high alarm and front window closed then the fan shut down.
●Fan speed is adjustable according to need to adjust the appropriate amount of fan.

STABLEFLOW™ Stable fan speed system

●High air flow fan, the system will automatically track and adjust the air volume and pressure, to balance the role of compensation, with the design of micro-shaped mesh layer outflow wind deflector, so that more air flow stability and uniformity.

Excellent imported sensor: Air Velocity Sensor and Pressure Plenum Sensor

●Imported air velocity sensors collect data of high precision, the performance of corrosion-resistant, anti-interference ability.
●Real-time monitoring and control cabinet status.
●Imported pressure plenum sensor responsive, effective monitoring the pressure of the work zone and plenum.

Ergonomically design

●Ergonomically angled front. Increase operating comfort, reduce work fatigue.
●Equipped with splash electrical outlet. The Weight-Balance front window can move up and down freely.
●INSIDE lighting system to provide the best illumination angle and brightness, reducing visual fatigue.
●Front windows with explosion-proof, anti-ultraviolet tempered glass, to better protect the safety of the operator.

Modern manufacturing processes

●Sheet metal parts use laser cutting and CNC bending technology. Cold-rolled sheets use three line of acidification anti-rust technology. Incubator surface use workmanship of spraying plastics.

Easy to clean

●Working surface with anti-corrosion wear of SUS304 stainless steel plate forming a seamless, leak-proof, easy cleaning use.
●Under the working surface with the drain valve, easy to clean and emissions.

Convenient Maintenance

●Removable combination tripod, easy to transport and move, Stand height adjustable.
●Eversion design of the front control panel to facilitate the electrical components, maintenance and replacement.
●In order to make the BSC able to achieve long-term safety requirements, it is recommended that the user to clean or replace the HEAP of the BSC in use for some time.

Technical Parameters:

Product ModelBIOCASION-A2
Biological safety cabinet
BIOCASION-B2
Biological safety cabinet

A2-1000G(D)A2-1300G(D)B2-1000G(D)B2-1300G(D)
clean ratingInternational standard ISO14644.1 CLASS 5(US209E,CLASS 100)
inflow velocity0.53m/s0.56m/s
down flow velocity0.35m/s
down flow velocity precision±0.015m/s
noise level(dB)≤65
vibration≤3μm(X、Y、Z direction)
HEPAefficiency 99.995%
light intensity(LX)≥650
pressure difference range (Pa)0~500
operate environmentenvironment temp.10-30℃,relative humidity  under 70%, no obvious vibration and dust
external dimensions(H*W*D) (mm)2060×1035×7402060×1340×7402120×1035×7402120×1340×740
dimension of working zone(H*W*D) (mm)670×965×560670×1270×560670×965×560670×1260×560
power supply                            Single-phase voltage source AC220V/50HZ
Power(W)300350300350
net weight(KG)175210175210
gross weight(KG)195235195235
package size (H*W*D) (mm)1650×1205×9101650×1510×9101710×1205×9101710×1510×910

more:http://en.kentonchina.com/biosafety-cabinet/20170531/16.html