Product Details

Universal Calibrator

Brand Name PI Controls UK

Product Features

Universal Calibrator

 

Testing & Maintenance

  • The Calibrator do not need any special maintenance other than periodic calibration check as mentioned earlier. For better performance, use the calibrator in a location that meets the following criteria

a.) Little or no vibrations

b.) No corrosive gases

c.) Minimal temperature fluctuation and near normal room temperature (25 C)

d.) No direct subject to Radiant heat

e.) Not in strong electromagnetic field

f.)  No excessive humidity

  • Before connectiing any wire to terminal strip, check following things

a.) Proper terminals are selected

b.) Appropriate parameter is selected. e.g. mV, mA Ohm, Temp.

c.) Select source or Sink Mode as per requirement

d.) In case of RTD, first read the parameter and then feed to the instrument under calibration.

e.) Whenever calibrator not in use, switch of the supply to enhance battery life.

f.) In case calibrator will not be used for longer duration, remove/disconnect batteries.

 

Testing Procedure

This procedure is required to confirm proper working of calibrator to avoid ambiguity.

Adjust: Select Switch to mV.

              Mode Switch to Source.

Adjust reading 100.00 on DPM with the help of  Coarse and Fine potentioner.

Confirm the output with standard traceable multimeter at mV terminals.

This must be within specified accuracy of ±0.1