Features
Three-step defibrillation process
Two-button operation
Extensive voice and visual prompts for the operator
Biphasic energy output
Lock-out protection to prevent inadvertent defibrillation
Automatic Self-test:Daily battery condition test& Monthly AED whole inter system test
Defibrillator
Output | Specification | |
Energy Sequence (user configurable) | Biphasic Truncated exponential | |
150, 150, 200J | ||
Charge Time from "Shock Advised | 10 sec. to 150J 15 sec. to 200 J | |
Analysis Time | 9 sec. | |
From start of analysis to shock ready: | from power up | Less than 20 seconds |
after 6 shocks | Less than 25 seconds | |
Audible Prompts | 22 audible prompts | |
Visual Prompts | LED prompts | |
Controls | Two buttons - On/Off, Shock | |
Output Energy Accuracy | ±15% into any impedance from 25 to 175Ω | |
20Ω to 200Ω |
Physical
Category | Specification |
Dimensions | 303 x 216 x 89 mm |
Weight | 2.0 kg |
Operating Temperature | 0ºC to 40ºC |
Operating Humidity | Relative humidity between 30% and 95% (non-condensing) |
Storage Temperature (without battery) | -20ºC to 55ºC |
Storage Humidity (without battery) | Up to 93% (non-condensing) |
Electrical Isolation
Category | Specification |
Power | Unit operates on internal battery only |
External Electrical Connections | No external devices are attached to the unit |
Risk Current Category | Internally powered equipment with defibrillator-proof BF type patient applied part (as per definition of IEC 60601-1 standard) |
Battery
Category | Specification |
Part No | CR123A-4x2 |
Non-rechargeable | 12V d.c. 2.8Ah |
Capacity | 100 discharges at 200 Joules or 120 discharges at 150 Joules |
Shelf Life (25ºC±15ºC) | 5 years |