Product Details

Single Channel Syringe Pump

Type Other
Brand Name LENCEN
Model Number BYZ-810
Place of Origin Jiangsu, China

Product Features

Specifications

Single channel syringe pump;
Three working modes:
Unique Human Voice Alarm system and accurate injection

Single Channel Syringe Pump  

Single Channel Syringe Pump

 

Detail:

1

1.Freely Stackable: Users can freely stack one syringe pump onto another to provide multiple Solutions, which have a wide range of clinical applications.
2.Transfer the message between the Nurse-call System and the pumps.
3.Three working modes: rate mode, time-volume mode, dosage-weight mode.
4.Unique Human Voice Alarm system and accurate injection rate make the injection process safer and more reliable.
5.Have store many brands syringes,Automatically calibration and compatible with any brand of syringes.
6.Large LCD screen displays working status.
7.KVO and Bolus functions.

 

 

Technical parameters of BYZ-810 syringe pump:

 

Unique Human Voice Alarm system

The device triggered the human voice prompt and stop injection automatically when there is any malfunction, making the infusion process safer and more reliable.

Rate:

50/60ml Syringe:  0.1ml/h~999.9ml/h   (0.1ml/h step)
1,000ml/h
~
1,500ml/h   (1ml/h  step)
30ml Syringe: 0.1ml/h
~
900.0ml/h    (0.1ml/h step)
20ml Syringe:  0.1ml/h
~
600.0ml/h   (0.1ml/h step)
10ml Syringe:  0.1ml/h
~300.0ml/h   (0.1ml/h step)

Flow Rate Accuracy:

Within ±3% (after correct calibration)

Mechanical precision:

Within ±2%

Bolus Rate:

50/60ml Syringe:  1,200ml/h
30 ml Syringe:  720ml/h
20 ml Syringe:  480ml/h
10 ml Syringe:  240ml/h

Purge Rate:

50/60ml Syringe:  1,500ml/h
30ml Syringe:  900ml/h
20ml Syringe:  600ml/h
10ml Syringe:  300ml/h

Volume Limit:

0.1ml~999.9ml

Total Injection
Volume:

0.1ml~9999.9ml (0.1ml step)

Occlusion:

High:   800mmHg ±200mmHg     (106.7kPa±26.7kPa
Medium: 50/600mmHg ±100mmHg  (66.7kPa±13.3kPa
Low:    300mmHg ±100mmHg    (40.7kPa±13.3kPa )

Alarms:

Injection soon finish, end of injection,occlusion, improperly installation of syringe, wrong setting, low battery, syringe loose etc.

Power Source:

AC 100V~240V, 50/60Hz; Internal rechargeable Li battery, capacity1,600mAh, 4 hours internal battery backup

KVO rate:

1ml/h

Fuse:

F1AL/250/60V, 2pcs inside

Power Consumption:

30VA

IP Classification:

IPX4

Equipment Classification:

Class I, internal power supply, Type CF

Operating Condition:

Ambient temperature: +5°C~+40°C;

Relative humidity:

20~90%

Transport & Storage Condition:

Ambient temperature: -30°C~+55°C

Relative humidity:

95%

Dimension:

280mm (L) × 210mm (W) × 130mm (H)

Weight:

2.2kg (net weight)

 

 

 