Specifications
The first ICU ventilator with CE mark in China .ICU ventilator with Air Compressor
multi ventilation wroking model
Aeonmed - The Operating Expert of OR & ICU
Features
The first ICU ventilator with CE mark in China
Pneumatically driven and electronically control for pediatric and adult
Detachable 10.4”TFT color screen decrease cross-infection effectively
Easy setting by navigator wheel knob & touch key
Three-level light and sound alarm, with concise words describing problems
Sensitive pressure trigger and flow trigger with base flow
Lung mechanics monitoring and loops
Emergency start function
Synchronized nebulizer functionresistant
Built-in battery
Modes of Ventilation
Volume Controlled (VCV), Assist/Control
Pressure Controlled (PCV) , Assist/Control
Synchronized Intermittent Mandatory Ventilation (SIMV)
Volume-controlled breaths (V-SIMV)
Pressure- controlled breaths (P-SIMV)
Pressure Support (PSV)
Spontaneous ventilation (SPONT)
Continuous Positive Airway Pressure (CPAP)
Combined: VCV + SIGH, SIMV (VCV)+PSV, SIMV (PCV)+PSV
Parameters
Tidal volume:0~1500ml
Respiratory rate:4~100bpm(used in VCV & PCV)
1~40bpm (used in V-SIMV & P-SIMV)
Inspiration time: 0.1~12s
Inspiratory pause time:0~4s
FiO2:21%~100%
Trigger sensitivity: Pressure (-2~0kPa, above PEEP)
Flow (0.5~20LPM, with base flow)
PEEP:0~4kPa
Psupport:0~6kPa
Pcontrol:0.5~6kPa
Nebulizer:30 minutes
Emergency start: Yes
Key lock: Yes
Monitoring
Pressure values: Paw, Pplat, Pmean, PEEP
Volume/Flow values: VTI, VTE, MV, MVspont
Time values: Ftotal, Fspont, I:E
Oxygen monitoring: Without O2 sensor
Real time curves: Pressure-Time, Flow-Time, Volume-Time, Pressure-Volume loop, Flow-Volume loop
l Respiratory mechanics, dynamic and static compliance, resistance, auto PEEP
Alarm
MV high, MV low, Circuit disconnected, Paw high, Paw low, Airway pressure continue high, VTE high, Fspont high, Tapnea, FiO2 low, Power supply failure, Nebulizer on, Battery low, Battery exhausted, Air supply down, O2 supply down, Air & O2 supply down
Technical data
Screen:10.4”TFT color screen (Detachable)
Gas supply:O2 ,Air (All gas must be medical level) , 0.28~0.6MPa
Power supply:AC110~240V, 50Hz/60Hz, 65VA
Maximum security pressure: 8kPa
Compliance:4mL/100Pa
Noise: 65dB(A)
Communication interfaces:RS232 port, VGA port
Dimensions (H×W×D): 400×303×250 mm
Weight:15kg
Environment requirements
Temperature: | 5°C~40°C(Operation) |
| -10°C~55°C (Storage) |
Relative humidity: | ≤80%, non-condensing(Operation) |
| ≤93%, non-condensing (Storage) |
Atmospheric Pressure: | 86~106kPa(Operation) |
| 86~106kPa(Storage) |
Altitude Operation: | 500~800mmHg / 3565~ -440m (Operation) |
| 375~800mmHg / 5860~ -440m (Storage) |