CE Approved Dental X-ray Tube for Replacement TOSHIBA X ray film machine tubes 70kV , O.4 Focal Spot

Brand Name DONGZI
Place of Origin China
Model Number KL11-0.4-70

Product Features

Delivery Time:30DAYS
Certification:CE SFDA

CE Approved High Frequency Dental X-ray Tube for Replacement TOSHIBA X ray film machine tubes 70kV O.4 Focal Spot

Quick Detail:

Type:Station anode x-ray tube

Application: For dental radiography unit

Model: KL11-0.4-70

Equivalent to TOSHIBA D-0411

Integrated high quality glass tube

Description:

KL11-0.4-70 Stationary anode X-Ray Tube is especially designed for high frequency intra-oral dental x-ray unit and available for a nominal tube voltage with DC.

KL11-0.4-70 tube has one focus.

The integrated high quality tube with glass design has one super imposed focal spot and a

reinforced anode. The high anode heat storage capacity ensures a wide range of applications for intra-oral dental application. A special designed anode enables an elevated heat dissipation rate which leads to a higher patient throughput and a longer product life. A constant high dose yield during the entire tube life is ensured by the high density tungsten target. Ease of integration into system products is facilitated by extensive technical support.

Applications:

KL11-0.4-70 Stationary anode X-Ray Tube is especially designed for high frequency intra-oral dental x-ray unit and available for a nominal tube voltage with DC.

Specifications:

Maximum Tube Voltage……………………………………..….………..…….……………70kV

MaximumTube Current……………………………..….……….………………………...…9mA

Maximum power (at 1.0s)………………………………………….……….……..............430 W

Maximum anode cooling Rate……………………………………………………………110W

Max.Anode Heat Content………….……….…………………..……………………...…...4.3kJ

Filament Characteristics……………………………………………...Ifmax=3.0A,Uf=3.2±0.5V

Focal spot ..…………………………………………………….……….….0.4(IEC 60336 2005)

Target Angle……………………………………………………………….………………..…12°

Target Material………………………………………………………………….………..Tungsten

Cathode type……………………….…………………………………………….…….W filament

Permanent Filtration………..…………..……………...Min. 0.5mmAl/50 kV(IEC60522/1999)

Dimensions………………………………..…………….……67mm length by 30mm diameter

Weight…………………………………………………………………………….…….100 grams

Competitive Advantage:

Available for the DC generator

Constant high dose yield

Excellent lifetime

You May Like

Find Similar Products By Category

You May Like

View More