Products Description
Faetures:
①. 150 Sample positions.
②. 160 Reaction cuvettes.
③. 180 Refrigerated reagent positions.
④. Optical filter/fiber bragg grating selectable.
⑤. 800 Tests per hour 1200 Tests with ISE module.
⑥. Built-in sample barcode system optional.
⑦.Probe with anti-collision function, liquid level detection function.
②. 160 Reaction cuvettes.
③. 180 Refrigerated reagent positions.
④. Optical filter/fiber bragg grating selectable.
⑤. 800 Tests per hour 1200 Tests with ISE module.
⑥. Built-in sample barcode system optional.
⑦.Probe with anti-collision function, liquid level detection function.
Product Paramenters
1.Reaction Tray
37±0.1℃,water bath
2.Mixer Probe
Dual mixer, auto frequency adjustment
3.Reagent Tray
2~8℃ cooling for 24 hours
4.Sample Probe
Liquid level sensor function Anti collision function Reagent volume real-time detection
5.Software
User-friendly software
6.Washing Probe
Dual design, high efficiency
37±0.1℃,water bath
2.Mixer Probe
Dual mixer, auto frequency adjustment
3.Reagent Tray
2~8℃ cooling for 24 hours
4.Sample Probe
Liquid level sensor function Anti collision function Reagent volume real-time detection
5.Software
User-friendly software
6.Washing Probe
Dual design, high efficiency
Model | MF-1200 | ||
Overall Performance | Throughput | 800Tests/hour, 1200Tests/hour with ISE(Optional) | |
Analysis Method | End-point, Fixed-time, Rate(Kinetic), Turbidimetry | ||
Sample & Reagent System | Sample Positions | 150sample positions | |
Reagent Position | 180reagent positions | ||
Sample Volume | 2~70μL | ||
Reagent Volume | 20~350μL | ||
Probe | Teflon coating,withanti-collision function,Liquid level detectionfunction | ||
Probe Washing | Automatic washing interior and exterior | ||
Reagent Cooling | Refrigerated tray with independent switch | ||
Reaction System | Temperature Control | Water bath37±0.1℃, real-time monitoring | |
Cuvettes | 160reusable cuvettes, optical length 6mm | ||
Mixer Probe | Independent stirring | ||
Washing | Automatic cuvettes washing | ||
STAT function | YES | ||
Optical System | Light Source | 12V/20W halogen lamp/fiber bragg grating optional | |
Spectrophotometry | Post-spectral spectrophotometry | ||
Wavelength | 12 wavelength from 340~800nm | ||
Absorbance | -0.3~3.0Abs | ||
Calibration & QC | Calibration | Linear:K factor, 1-point ,2-point and multipoint linear Non-Linear:Spline,Polygon, Index, Ogarithm, Logit-4P, Logit-5P | |
Quality Control | Real-time QC, Westgard multi rule, Cumulative Sum Check,Twin Plot(2D) | ||
Data Management | Software | Windows 7/8/10 , 32or 64bit | |
LIS System | Bi-direction,support HL7 protocol | ||
Interface | LAN port access | ||
Printer | External,multiple reporting mode available | ||
Working Conditions | Power Supply | AC220V±10% 50/60Hz, 110V±10% 60Hz, 1500W | |
Temperature | 10~30℃ | ||
Water Consumption | Deionized water:20L/H | ||
Humidity | 30-80% |